KE-50 Datasheet, Funktion
GS Oxygen
Features
Applications
Sensors
GS oxygen sensors, model KE-series, are
newly developed galvanic cell type
sensors. By using acid electrolyte, the
sensor has an exceptionally long life and
excellent chemical durability.
* Long life
KE-25 - 5 years
KE-50 - 10 years
*Virtually no influence from CO2, CO,
H2S, NOx, H2
* Low cost
* Operates in normal ambient
temperatures
* Stable output signal
* No external power supply required for
sensor operation
* No warmup time is required
Medical:
Anesthetic instruments,
respirators, oxygen-
enrichers
Biotechnology: Oxygen incubators
Food industry: Refrigeration,
greenhouses
Safety:
Air conditioners,
oxygen detectors,
fire detectors
Structure of KE-25 / KE-50
Lead wire (anode)
Lead wire (cathode)
Anode (lead)
O-ring
Plasticdisc
Plastic ring
Black
Red
Resistor
Thermistor
Acid electrolyte
Sponge disc
Cathode (gold film)
Teflon membrane
Principle
Specifications
GS oxygen sensors incorporate a lead oxygen
cell with a lead anode and a gold cathode,
using a specific acid electrolyte. Oxygen
molecules diffuse through a non-porous
fluorine resin membrane into the
electrochemical cell and are reduced at the
gold electrode. The current flowing between
the electrodes is proportional to the oxygen
concentration in the gas mixture to be
measured. The signals are measured as
terminal voltages of the resistor and the
thermistor for temperature compensation.
The change in output voltages represents
the oxygen concentration.
Model
KE-25
KE-50
Measurement range
Accuracy
Operating temperature
0~100% O2
±1% (full scale)
5~40˚C
0~100% O2
±2% (full scale)
5~40˚C
Storage temperature
-20~+60˚C
-20~+60˚C
Response time (90%)
14±2 seconds
60±5 seconds
Initial output voltage
(at 21%O2 and std test conditions
of 25˚±1˚C, 60±5%RH, 1013±5hPa)
10.0~15.5mV
47~65mV
Life expectancy
approx. 900,000% hours approx. 1,800,000% hours
(approx. 5 yrs. in 20˚C air) (approx. 10 yrs. in 20˚C air)
Remarks: Life expectancy is specified as % hours (O2 concentration % x hours)





SeitenGesamt 2 Seiten
PDF Download[ KE-50 Schematic.PDF ]

Link teilen




Besondere Datenblatt

TeilenummerBeschreibungHersteller
KE-50GS Oxygen SensorsETC
ETC

TeilenummerBeschreibungHersteller
CD40175BC

Hex D-Type Flip-Flop / Quad D-Type Flip-Flop.

Fairchild Semiconductor
Fairchild Semiconductor
KTD1146

EPITAXIAL PLANAR NPN TRANSISTOR.

KEC
KEC


www.Datenblatt-PDF.com       |      2020       |      Kontakt     |      Suche